Appendix 1: Activated carbon unit
© Elma Schmidbauer GmbH
BA_Elmasolvex_VA_EN_Vers.05.2018
105
14
Appendix 1: Activated carbon unit
14.1
Description
The activated carbon unit consists of the base frame, the
retaining bracket and 1 condensate separator and 2 activated
carbon cartridges.
The base frame can be either floor-standing or wall-mounted.
The condensed solvent vapours are collected in the condensate
separator and must be drained regularly by the user. Suitable
activated carbon granulate must be filled in the two activated
carbon cartridges and changed regularly by the user.
Activated carbon unit, 2 fillings of activated carbon granulate,
each of 500 g, funnel for filling the activated carbon cartridges,
1 set of PE filter inserts, labels for noting the maintenance,
PTFE sealing tape, fastening material for optional wall mounting
of the activated carbon unit.
Fire and explosion hazard!
Solvents are separated at the activated carbon unit.
The instructions from
Chapter 8
of this operating manual are
applicable.
Observe the applicable safety regulations for handling
solvents.
Keep all kinds of ignition sources away.
Prevent ignition sparks from electrostatic discharge. Discharge
possible electrostatic charges (body charge) by touching any
grounded equipment before handling flammable media: e.g.
water tap, metal surface of the case of the cleaning machine or
use ESD protective equipment (ESD arm band).
The connected cleaning machine must not be operated during
commissioning and maintenance of the activated carbon unit.
Used operating materials must be disposed of in accordance
with the regulations. No disposal via the sewer system! Dispose
of used operating materials in accordance with the national
disposal regulations for the media (see safety data sheet).
Used activated carbon granulate is also taken back by the
supplier or its manufacturer.
Scope of delivery
DANGER
Disposal of used
operating materials