February 2022
C6057701-01
Q-Series Quick Guide
Please download the full instruction manual from the Ultrawave website
Electrical Safety instructions
The Q-Series Ultrasonic Baths are designed to be connected to a 230VAC fully
earthed electrical supply via a three-pin plug.
It can be dangerous to operate the Ultrasonic Bath without an earth
connected.
The Q-Series machines are rated at up to 1050W maximum.
The detachable mains lead is rated at 250V 6A.
NEVER FIT A LEAD WITH A
LOWER RATING.
The mains plug is fitted with a 5A fuse.
NEVER FIT A FUSE OF A HIGHER
RATING.
The IEC electrical socket on the rear of the unit is fitted with 2x T5A (time-
lag) fuses.
NEVER FIT A FUSE OF A HIGHER RATING.
The detachable mains lead is the disconnect device and should remain
accessible while the machine is in use.
Ensure that any excess mains cable is stored neatly.
General Safety Instructions
Do not place fingers, hands, or other parts of the body in the liquid in the bath
when it is sonicating.
Always ensure the liquid is above the level sensor when in operation.
Do not place parts or containers directly on the bottom of the cleaning tank.
Always use a basket or instrument cassette to support items to be cleaned.
Not doing so may damage the bath and invalidate your warranty.
Never use toxic, flammable, acidic, caustic, or corrosive chemicals or liquids
not recommended by Ultrawave.
Ensure the electrical supply is disconnected before moving/emptying.
Before cleaning, drain the bath. Clean by wiping with a clean non-abrasive
cloth.
The user should familiarise themselves with the
full instruction manual
before
operating the equipment. Safety may be impaired if they are not followed.
Ultrawave will not be responsible for damage or injury caused by incorrect use
of the equipment.
The user should apply to Ultrawave or its agent for advice on cleaning
techniques or detergents.