Operating faults
© Elma Schmidbauer GmbH
BA_Elmasolvex_VA_EN_Vers.05.2018
85
Fault message
Fault description
Fault rectification
Error No. 51 - 82
Abortion
Critical fault!
Explosion safety not
guaranteed.
Program has not been
ended.
Tripping of the vacuum
safety switch.
Cleaning chamber too long
without vacuum.
Medium possibly drains too
slowly.
Check the following possibilities:
•
Air intake and exhaust openings in the
cleaning chamber (
Fig. 3.12.1.A
)
clogged?
•
Condensate separator, activated
carbon filter (if connected) clogged?
•
Exhaust air tube (
chapter 4.3
) kinked?
•
Sieve insert in the cleaning chamber
clogged (
Fig.9.1.4.1
)?
•
Suction filter in the media tank (
Fig.
3.13.1.G
) clogged / blocked?
•
Working level fill level sensor
(
Fig.9.1.4.1.A
) soiled? Clean with
suitable cloth if necessary.
•
Work level fill level sensor
(
Fig.9.1.4.1.A
) damaged (scratches,
cracks)?
•
Quick couplings on the media tanks
(
Fig. 3.13.1. B/C
) correctly engaged?
•
Can the valves of the quick couplings
(
Fig. 4.4.1.2.J
) be pressed in / moved
correctly?
•
Can the valves of the quick couplings
on the media tanks (
Fig. 4.4.1.1.F
) be
pressed in / moved correctly?
•
Is the cover of the cleaning chamber
closed correctly?
•
Cover seal of the cleaning chamber
(
chapter 10.1.1.B)
damaged / slipped?
•
Emergency ventilation in the cover of
the cleaning chamber (
chapter
11.3.2.A
) leaking?
•
Switch off cleaning machine and
restart: Contact Service if the fault
persists.
Error No. 83-114
Abortion
Critical fault!
Cleaning chamber
overfilled
Program has not been
ended.
Tripping of the overfill level
fill level sensor
or
Fill level sensor for overfill
level tripped more than three
times, however only for a
short time, (multiple
splashes) during any process
step
Check the following possibilities:
•
Cleaning basket inserted correctly?
•
Sieve insert (
Fig.9.1.4.1
) in the
cleaning chamber inserted correctly?
•
Fill level in the media tank is too low
(and air is drawn into the working
chamber as a result)?
•
Working level fill level sensor and
overfill level fill level sensor
(
Fig.9.1.4.1 A/B
) soiled? Clean with
suitable cloth if necessary.
•
Working level fill level sensor and
overfill level fill level sensor
(
Fig.9.1.4.1 A/B
) damaged (scratches,
cracks)?