60
60
Section 5: Preventative Maintenance and Safety
5. Repeat the spray treatment every 2 or 3 minutes for a total of 20 minutes. Do not allow cleaning
solutions to dry on the instrument surfaces.
6. Rinse the inner bowl and lid thoroughly with water.
7. Spray and wipe the exterior surfaces with decontamination solution such as REF: SS-133.
CAUTION:
Do not flood the display panel with excessive moisture. Any moisture that seeps through could
damage the internal electronics.
8. Repeat the spray treatment of exterior surfaces every 2 or 3 minutes for a total of 20 minutes. Do not
allow cleaning/decontamination solutions to dry on the instrument surfaces.
9. Wipe surfaces thoroughly with a cloth soaked in water until all decontamination solution has been
removed.
10. Immerse or generously spray the carousel and lid with decontaminating solution. Allow the solution to
react for 20 minutes.
11. Thoroughly rinse the carousel and lid with deionized or distilled water.