43
43
Section 3: Controlling and Customizing Stainer Functions
Manually Entering Specimen Information
With Stain Slide Tracking and Manual Entry enabled in the QC Maintenance menu:
1. Press
Load Slides
on the Main menu.
2. Press
Add
to reveal the keypad.
3. Enter slide information (maximum of 24 characters) and press
Enter
.
4. To change or delete the entry, select the entry on the display and press
Edit
or
Remove
.
5. Load slides and run stain cycle as shown in Section 4.1.
Manually Entering Reagent Information
1. Press
Reagents
from the System Setup menu or press the reagent status icon on the Main menu to
reveal the Reagent Information menu.
2. Select the desired reagent and press
Change
.
3.
Press the desired field (Reagent REF, Expiration Date, or Lot Number (Serviced Date/Time
automatically gets entered); enter the information on the keypad and press
Enter
.
NOTE:
Reagent REF number must be a valid ELITechGroup REF number for the selected reagent.
Incorrect entries will generate an error message.
4. When all the information has been entered, press
Save
.