ELGA
PURELAB Classic
Operator Manual
PURELAB Classic
Version 2 11/15
Page 9
3.3 Technical Specification
The Technical Specifications for the
PURELAB Classic
are as
follows:
Feedwater
*
Parameter
Limits
Source - Originally from potable supply,
then pre-treated
Preferably reverse osmosis (RO) or filtered service deionisation (SDI) or distilled.
Note:
mixed bed or twin bed deionised supplies should be cation limited at exhaustion
.
Fouling Index (max)
1 for all models. A 0.2 micron membrane prefilter is recommended for all non-RO feeds.
Service Deionisation (SDI) - M
-cm
1 M
-cm minimum resistivity at exhaustion.
Reverse Osmosis (RO) - µS/cm
Recommended < 30 µS/cm
Free Chlorine
0.05 ppm max.
TOC
Recommended 50 ppb max.
Carbon Dioxide
30 ppm max.
Silica
2 ppm max.
Particulates
Filtration down to 0.2 micron advisable to protect internal and/or point of use filters.
Temperature
1 - 40°C - Recommended 10 - 15°C
Flowrate
(maximum requirement)
130 l/hr
Drain requirements (gravity fall with air gap).
Maximum during service
Up to 2 l/min
Feedwater Pressure
0.7 bar (10 psi), maximum, 0.07 bar (1 psi) minimum
* Contact technical support for advice on feedwaters outside the range listed.
Note:
Different system configurations are available for different feedwater sources. See system set up.
Dimensions
Height
490 mm (19.3")
Width
410 mm (16.2")
Depth
365 mm (14.4")
Weight
DI
14.0 kg (30.8 lb)
UV
14.5 kg (32.0 lb)
UF
14.5 kg (32.0 lb)
UVF
15.0 kg (33.1 lb)
Connections
Inlet-quick connect
8 mm (5/16") OD
Flush - UF/UVF
8 mm (5/16") OD
Positioning
Wall, bench or under bench mounted.
Environment
Clean dry indoor. Temp 5 - 40°C.
Humidity max 80% non-condensing.