7. MAINTENANCE
Any maintenance work not detailed in this handbook should be carried
out by an approved supplier or distributor. If further information is
required on any aspect of maintenance please contact Customer
Service.
WARNING!
ALWAYS CHECK THAT THE MAINS
ELECTRICAL
POWER
AND
FEED
WATER SUPPLIES ARE SWITCHED
OFF
BEFORE
ATTEMPTING
TO
CHANGE A CONSUMABLE ITEM.
7.1 Replacing the Purification Pack (LC186)
The Purification Pack should be replaced in the following
circumstances:
The quality monitor indicates that the pack requires
changing.
If the system is being e-commissioned or sanitized after an
extended period during which it was not used.
If the pack has exceeded 12 months use as indicated by the
consumable reminder alarm.
Step 1 - Switch Unit Off
1.
SWITCH OFF
PURELAB Classic
at the power inlet
module.
2.
ISOLATE inlet water supply.
3.
REMOVE POU filter, if fitted.
4.
OPEN the front door
.
Step 2 - Fit New Purification Pack
1. REMOVE a new Purification Pack from its packaging.
2. REMOVE the sealing plugs from inlet and outlet port.
3.
WET ‘O’ rings and SLIDE new Purification Pack into the
right hand position.
4. POSITION Purification Pack onto spigots.
5. PUSH into unit.
6. ENSURE that the Purification Pack is fully engaged and
dropped down past the pack retainers.
Purification
Pack
Location of Purification Pack
Top Cap
Removal of Purification Pack