ELGA
PURELAB Classic
Operator Manual
PURELAB Classic
Version 2 11/15
Page 39
Step 4 - Fit Sanitization Blocks
1. REMOVE sanitization block
– LA698 from the inside of
the door.
2.
WET ‘O’ rings on both sanitization blocks.
3. SLIDE the LA698-US into unit the unit in the left-hand
side (lower position).
4. SLIDE the LA698 into the right-hand side (upper
position).
5. SWITCH ON power.
8.
Note: The
PURELAB Classic
will automatically
sense that the bypasses are fitted and default to
the sanitization cycle.
6. ACCEPT starting the sanitization cycle.
The
PURELAB Classic
will start to recirculate for 10
minutes.
Step 5
– Contact Time
1. PRESS the process button once 05:00 or less is
displayed on the screen
2. ALLOW the unit to stand for a minimum of 36 minutes,
a maximum of 60 minutes
3. PRESS the process button.
The unit will then proceed to recirculate for the
remainder of the time and then automatically proceed to
the rinse procedure. The unit will alternate between a
rinse to drain and recirculate for a total of 44 minutes.
Upon completion the alarm will sound for 4 seconds.
The sanitization cycle is now complete and the unit will
now display the next sanitization reminder alarm date.
Step 6 - Remove Sanitization Blocks
1. PUSH and TILT by-pass blocks.
2. SLIDE out of unit.
Step 7 - Replace Purification Packs
1. INSERT Purification Pack into position.
2. Allow the
PURELAB Classic
to recognize the
Purification Pack and ensure that it has been replaced
into the correct position.
3. INSERT Purification Pack into right hand side of the
compartment (upper position).
4. Allow the
PURELAB Classic
to recognize the
Purification Pack and ensure that it has been replaced
into the correct position.
5. ENSURE Purification Packs are locked into position.
6. CLOSE front door.
Container
Level Indicator
Spigots
Fit sanitization Block
Slide out
Tilt
Push
Bypass block
Hangers
Remove sanitization Block