PURELAB Classic
Operator Manual
ELGA
Page 20
PURELAB Classic
Version 2 11/15
Step 9 - Temperature Alarm Setting
The temperature of the water is constantly monitored to ensure
that it does not increase to an unacceptable level. The
PURELAB Classic
will alarm and switch the process off if the
alarm point is exceeded.
1. PRESS SCROLL
button to highlight appropriate box.
Note:
To operate the unit without a temperature
alarm PRESS SCROLL
until no boxes are
highlighted. The unit will switch the process
mode off at 50°C to prevent damage.
2. PRESS TICK
button.
Step 10 - Display Viewing Angle Adjustment
The angle of the display can be electronically adjusted up and
down to optimize the display graphics visibility.
1. ROTATE Dispense Controller slowly to adjust viewing
angle.
2. PRESS TICK
button.
Step 11 - Auto-Restart
This allows the selection of the AUTO/MANUAL restart option. If
auto restart is selected the unit will automatically restart after a
temporary loss of power to the unit. In manual mode the unit will
remain in standby mode.
1. PRESS SCROLL
button to highlight appropriate box.
2. PRESS TICK
button.
Step 12 - Reservoir Level
When operating the system from a reservoir it is recommended
that a low-level switch be connected to protect the system from
running dry.
1. PRESS SCROLL
button to highlight box, (if reservoir
and level control are attached).
2. PRESS TICK
button.
The Controller Set-up is now complete.
Note:
To escape from any of the set-up screens
press the PROCESS button.
Display Viewing Angle Adjustment
Auto-Restart Screen
Reservoir Level Screen
Temperature Alarm Setting Screen