ELGA
PURELAB Classic
Operator Manual
PURELAB Classic
Version 2 11/15
Page 21
5.5 Setting Up Password / Resetting Replacement Timers
CAUTION!
Before resetting any of the Consumable
Replacement/Reminder Dates, ensure
that the appropriate new Consumable
has been installed and correctly located
in the
PURELAB Classic.
Step 1 - Enter Consumable Replacement Timer Set-up
1. SWITCH OFF at power inlet module.
2. PRESS and HOLD left hand button and SWITCH unit
back on. The unit will now enter the Consumable Timer
set up display.
3. RELEASE left hand button.
Step 2 - Password Change
Note:
If you do not wish to change password then
proceed to instruction 4.
1. ROTATE Dispense Controller to enter password code.
The default password is 000.
2. PRESS Dispense Controller once.
3. ROTATE Dispense Controller to enter new password.
4. PRESS TICK
button to accept new password.
Step 3 - Purification Pack Replacement Date
The replacement date for Purification Packs is automatically
calculated from information held on data tags secured to the
pack. The information displayed confirms the replacement
dates and the serial numbers of the packs that have been
installed.
Note:
At this stage the unit will not have been fitted
with any Purification Packs and the date
displayed
will
not
be
relevant.
Upon
installation of the packs this date will be
changed and reconfirmed.
1. PRESS TICK
button.
Step 4 - UV Lamp Replacement Date (if fitted)
1. PRESS RESET
button to reset UV Replacement
Date
OR
PRESS TICK
button to accept Replacement Date
and proceed to Filter Replacement.
2. PRESS TICK
button to confirm that resetting is
required
OR
PRESS CROSS
button
to abort reset.
3. PRESS TICK
button.
UV Lamp Replacement Date Screens
Password Screen