24
Elcom, spoločnosť s ručením obmedzeným, Prešov
Elcom, spoločnosť s ručením obmedzeným, Prešov
25
Uniq 150
Service
manual
Service manual
Uniq 150
It is possible to set hinge tilt after its removing and setting marked screws (stoppers) and actually
change Uniq 150 execution. It is necessary to adjust lock hinge and change direct cuff to angled cuff when
change execution from standard to N3 or T1.
Table 2.9.1. – Tilt setting
Execution
N1
N2
N3
N4
N5
N6
T1
T2
T3
Recommended
way of placing
table/wall
table/wall
table/wall
table/wall
table/wall
table
ceiling/floor
ceiling/floor
ceiling/floor
Angle
–15°/15°
(75°/105°)
0°/30°
(90°/120°)
78°/107°
(–12°/17°)
9°/39°
(99°/129°)
24°/54°
(114°/144°)
90°/117°
(0°/27°)
–25°/5°
(155°/185°)
87°/117°
(–93°/–63°)
102°/132°
(–78°/–48°)
Stopper position
B-B
A-A
B-B
B-B
A-A
A-A
B-B
B-B
A-A
Lock hinge
P: Y-Y
P: X-X
U: Z-Z
P: Y-Y
P: X-X
U: V-V
U: Z-Z
P: Y-Y
P: X-X
Cuff
Direct (P)/Angled (U)
P
P
U
P
P
U
U
P
P
Stand
Leg (Lx)/Tube Holder (TH)
L3
L3
L2
L1
L1
L3
TH
TH
TH
L1 – cables gap direction:
forward
; leg jambs direction:
forward
L2 – cables gap direction:
forward
; leg jambs direction:
backward
L3 – cables gap direction:
backward
; leg jambs direction:
outward
Figure 2.9.3. – Hinge structure and tilt setting
B
B
A
A
Y
X
P
T1
T2
T3
N5
N4
N3
N2
N1
Examples of positions:
N6
Lock hinge
Hinge
B
B
A
A
V
U
Z
2.10. MS reader installation
1.
It is necessary to adjust these components according to figure:
a.
Rear cover of Uniq 150 – drill a hole
b.
Side profile – cutting according to reader holder
c.
Right stripe of front cover – cutting (only Nx)
d.
Left stripe of front cover – cutting (only Tx)
2.
Unscrew particular screws according to execution (Nx/Tx).
3.
Cut out cable connector of MS reader and insulate the cable ends.
4.
Reader cable put through the side profile (
ø
4), then drilled hole of rear cover and finish adjustment of MS
reader cable:
Содержание Uniq 150
Страница 1: ...Uniq 150 Uniq 150 Service manual version 1 4 Elcom spolo nos s ru en m obmedzen m Pre ov...
Страница 5: ...Elcom spolo nos s ru en m obmedzen m Pre ov 5 Service manual Uniq 150 1 General specification...
Страница 7: ...Elcom spolo nos s ru en m obmedzen m Pre ov 7 Service manual Uniq 150 2 Structural part...