_NV Ron 2000 4.2
17/37
3/25/2016
The screen will read
OK
and then
MONTH
. Press
ENTER
and the display will
show four digits with the two leftmost digits blinking. This is the month. Use
the arrow keys again to select the month and press
ENTER
.
Continue with this process again for the day, hour, and minute. Press
ESC
to
finish setting the clock and exit back to the options menu. You may select
another option or press
ESC
again in order to exit from the options menu.
9.4 RS-232 Digital Output Communication Data (if
included)
Possible baud rates: 9,600 - 19,200 - 38,400 - 115,200.
Length: 8 bits
Stop bits: 1 bit
Paring bit: none
Data output modes:
The RS-232 output on Ron systems has two modes of data output that are
selectable by the user: on demand and continuous stream.
In on demand mode, data will be sent only when the operator uses the
function (see PRINT section). The continuous stream of data mode is
factory set to one of two versions that cannot be changed by the user. These
are:
1) unconditional constant flow
2) update on value change
The first will send a signal at a rate of about once per second when in use.
The second will send data on when there is a change in value detected by the
indicator.