_NV Ron 2000 4.2

14/37

3/25/2016

Press

ESC  +  ENTER

simultaneously  until  you  hear  the  beep,  then  quickly

press

TARE

.

The display will show the current units in use e.g.

Lbs

,

M.  TONS

etc. and

will then display

LOAD.0

.  At this point all weight/force must be removed

from the system.
Once the system is unloaded, press

ENTER

.  The display will flash

WAIT

for

a few seconds and then

L.VALU

(load value).  This means that the load of a

known weight intended for calibration is ready to be applied.

Once the known weight has been applied (lifted), press

ENTER

. The display

will show

APPLY

followed by the max capacity of the system. Using the

arrow keys ↑↓, set the load value you intend to use for the

calibration.  If no key is pressed for 8 seconds, the system

will exit out of calibration mode and you will have to start

again  from  the  beginning.    Once  the  load  value  on  the

display  matches  that  of  the known  weight currently  loaded,  press

ENTER

.

The  display  will  flash

WAIT

and  then

OK

.    This  means  the  system has

successfully been  adjusted  and the  display will  return  to  the  standard

measuring screen.

At any time, you may press

ESC

to exit out of the calibration process.  The

display will read

RETRN

and then revert back to the main measuring screen.

9. Options

Using the various available options for the Ron 2000

I

f your Ron system has addition optional features, you will need to use the

main code to access them.  To do this, press

ESC + ENTER

simultaneously.

The screen will display

CODE?

Using a short press, press

MAX

and then

press

ZERO

.

The screen will briefly show

MENU

, then one of the codes for an optional

function e.g.

TIME

,

UNITS

or

PRINT

.  If the code was not accepted the

display will read

FAIL

. Start again at the beginning of the code process.

Note:

By holding down

an arrow key, you can

gradually increase the rate

of scrolling.

Содержание RON 2000

Страница 1: ...RON 2000 DYNAMOMETER CRANE SCALE Eilon Engineering Weighing Systems Ltd Instruction Manual Version 4 2 Generation 4...

Страница 2: ...N M A N U A L F O R The Ron 2000 Shackle Type Hook Type Eilon Engineering Industrial Weighing Systems Ltd Phone 866 669 6122 Web www eilon engineering com E mail info eilon engineering com Shackles no...

Страница 3: ...9 3 Time setting the real time clock if included 16 9 4 RS 232 Digital Output Communication Data if included 17 9 5 Print Mode Selection only if the system is equipped with the RS 232 output 20 9 6 B...

Страница 4: ...angerous and is therefore STRICTLY PROHIBITED except during the system s annual safety testing This testing must be performed by qualified personnel and allows the system to be overloaded up to 25 of...

Страница 5: ...er becomes familiar with the particular needs of their system the need for calibration may vary Between calibrations the user can verify whether the systems are still calibrated correctly by using a k...

Страница 6: ...ns are observed No amount of safety features and engineering can be a substitute for common sense and a desire to work safely Ron systems are prohibited by the manufacturer and or seller to be used in...

Страница 7: ...ny attempt has been made to open any part of the system by any parties other than those specifically authorized by Eilon Engineering Ltd Batteries are not covered by this warranty The system supplied...

Страница 8: ...ometerbasicinformation The Ron 2000 is available either as a Shackle or Hook Type The system consists of a Load Cell and an indicator R R Ro o on n n 2 2 20 0 00 0 00 0 0 l l lo o oa a ad d d c c ce e...

Страница 9: ...ries rated 3AH will result in at least 2000 hours of battery life more than two months Occasional use will extend the battery life up to several years During use press ENTER to check the current batte...

Страница 10: ...r selectable 21 Wireless communication to additional display or PC 2 Operation Basicoperationof Ronsystemsandhelpfultipsandtricks Before operation be sure the shackles you intend to use are appropriat...

Страница 11: ...MAX key the MAX sign will be displayed The key should be pressed continuously until the beep is heard and visual sign appears Exceptions are turning the system on or off which does not require a long...

Страница 12: ...stem features a MAX aka PEAK HOLD function MAX will store the maximum weight that the system has detected since it was last powered on or since the Gross Net mode was changed Press the MAX key until t...

Страница 13: ...checked by an authorized inspector before being used again Please note that both overload warnings remain functional at the same levels 100 130 of full capacity whether the system is in Gross or Net m...

Страница 14: ...rom low voltage When the system is turn on the batteries energy level is displayed as a percentage e g 100 During use press the ENTER key to check the current battery level Optional Rechargeable Batte...

Страница 15: ...and may result in the batteries exploding If you are using disposable batteries in a system sold with rechargeable ones please remove the charger from the system s carrying case or place of storage i...

Страница 16: ...the best and most accurate results A known weight between 80 100 of capacity may be used but never a weight more than this If the system includes a dampened display averaging the dampening should be...

Страница 17: ...ain from the beginning Once the load value on the display matches that of the known weight currently loaded press ENTER The display will flash WAIT and then OK This means the system has successfully b...

Страница 18: ...ed Press ENTER again and the display will give the current level of set point 1 Use the arrow keys to level To scroll more quickly hold down the arrow key At first the rightmost digit will change Cont...

Страница 19: ...e the arrow keys again and scroll to the desired unit of measurement The available units are Lbs K NTON Kilo Newtons DECA N Deca Newtons N WTON Newtons KG M TON Metric tons S TON Short tons When the d...

Страница 20: ...cation Data if included Possible baud rates 9 600 19 200 38 400 115 200 Length 8 bits Stop bits 1 bit Paring bit none Data output modes The RS 232 output on Ron systems has two modes of data output th...

Страница 21: ...measurement Data 1 7 2 tab space tab 1 3 UNITS Measurement units KG S TON DECA N K NTON LBS LB 10 S TON M TON 2 5 6 6 3 6 5 5 4 tab space tab 1 5 TARE Exclusion of fixed weight Net Gross 3 5 6 tab spa...

Страница 22: ...0 M TON Net DATA 0 000 M TON Net DATA 0 000 M TON Net DATA 0 000 M TON Net DATA 0 315 M TON Gross DATA 0 315 M TON Gross DATA 0 315 M TON Gross DATA 1 015 M TON Gross DATA 1 010 M TON Gross DATA 1 715...

Страница 23: ...tant flow default the displayed valued will be output at a rate of about once per second 2 Continuous upon value change the displayed value will only be output when there is a change in the reading If...

Страница 24: ...y Value and total number of loads D L 1 reading 1 from Data Logger memory D L 2 reading 2 from Data Logger memory D L etc 5 DATE TIME if the system is equipped with the optional real time clock 9 6 Ba...

Страница 25: ...e system will reject it This filter is based on the fact that a load reading of zero must be sensed by the system in between loads If the system does not find a zero value before the current load is e...

Страница 26: ...obtain the load value in the new selected measuring units 9 9 Data Logger standard manual if included The standard data logger D L stores measurements in the indicator s internal memory Memories are...

Страница 27: ...nter C Seeing the total number of D L entries Press DATA L DATA L will appear on the screen Then press TARE The display will read DL N and then show the total number of readings currently stored in th...

Страница 28: ...hat affect the weight value stored MAX MIN AVERAGE MAX The maximum value registered by the system during each time interval will be logged in the memory To begin logging the user must switch from ADL...

Страница 29: ...ser to select the time interval T by selecting Days up to 31 and or Hours up to 24 and or Minutes up to 60 and or Seconds up to 60 Once the selection is completed press ESC to exit and return to the o...

Страница 30: ...of four wire rope falls the system will multiply the sensed load by four and display the result The system will multiply the sensed load by any number selected by the user to suit the actual reeving a...

Страница 31: ...alculates the average load based on the measurements taken over a given number of seconds here called T When measuring begins the first reading will show after about one second After two seconds the d...

Страница 32: ..._NV Ron 2000 4 2 29 37 3 25 2016 E RS 232 0 1V A V B GND for 0 1V C RS 232 D GND for RS 232 E RS 485 A NO B NO C A RS 485 D A RS 485 E GND...

Страница 33: ...then try turning it off If it does not switch off it can still be used until you have an opportunity to send it to an authorized service center With full batteries the system will function for more th...

Страница 34: ...and held indicator b The system should be checked in an authorized service lab 6 The system switches itself off a Check the batteries see battery care section b Try to operate the system without the e...

Страница 35: ...econds and Reset Data Logger S 011 Eeprom failure E 015 Active Real time Clock RTC power Check Replace battery 020 RS232 or SetPoint still active wait several seconds and try again S 25 26 Eeprom ON O...

Страница 36: ...ds size cap in short tons model G209A 1 3 8 31 3 M650A 2 3 8 31 3 M650A 3 31 3 M650A 5 5 8 3 4 7 M652A 10 1 11 8 15 M655A 12 1 11 8 15 M655A 15 11 8 11 4 18 M656A 20 13 8 G 2140 G 5263 25 11 2 11 2 30...

Страница 37: ...6 20 23 25 indicator 3 5 6 7 8 9 11 12 17 18 21 23 24 25 30 31 keyboard 3 6 8 29 laboratory 2 13 30 load cell 2 3 5 7 8 24 26 27 29 max 6 8 9 13 14 20 21 22 24 25 31 overload 4 10 27 plug 7 print 14 1...

Отзывы: