8. Calibration
Performing additional calibrationand adjustment to the system
Note
It is recommended that calibration be performed by authorized and skilled
personnel only! It should be performed using an accurate known weight or
in a laboratory.
U
nless otherwise required by local laws, it is recommended that the system
be calibrated annually. The weight used in the calibration must not exceed
the system's capacity.
BEST PRACTICE
It is recommended that calibration and adjustment are preformed
using a known weight that is 80% of the system's maximum
capacity. This will give the best and most accurate results. A known
weight between 80% – 100% of capacity may be used but never a weight more than
this.
If the system includes a dampened display/averaging, the dampening should
be shut off before calibrating (see Dampened Readings section). The system
is protected by two different codes to prevent calibration by unauthorized
personnel.
Main Code:
Press
ESC + ENTER
simultaneously until you hear the beep. The display will
show
CODE?
Quickly press
MAX
then
ZERO
. The display will flash
MENU
and then show one of the available options on the system, e.g.
UNITS
or
MESSG
(message) etc. This means you are now in the main
options menu.
Using the arrow keys ↑↓ scroll to
CALIB
and press
ENTER
. The display will
once again show
CODE?
prompting you to enter the secondary calibration
code.
Secondary Code: