18 |
P a g e
4-4. Display Screen
There are
2 screens that can be displayed.
1)
“FLOW RATE/ PRESS”
2)
“P.LIM/P.FREE/TIMER (Pressure Limit/Pulse Free mode/Timer)”.
<FLOW RATE/ PRESS> Displays the set flow rate and current pressure.
<P.LIM/P.FREE/TIMER> Displays the Pressure Limit /Pulse Free mode*/Timer)
*
Pulse free mode is a feature that reduces pressure pulses during pump operation. The pump senses
pressure fluctuation, and quickly learns to compensate to reduce pulsing. Please make sure to set this to
<PRESET> when doing analysis.
4-5. Log Function
There is also a log function built-in for recording the total liquid pumped from a given start date. It can be
used to record maintenance actions, such as installing a new pump seal or column. This log does not
have to be set, but we recommend that you use it to aid in troubleshooting and schedule regular
maintenance.
Switch display key
PUMP A LOG START DATE
PUMP B LOG START DATE
PUMP A FLOW RATE/PRESS.
PUMP B FLOW RATE/PRESS
.
PUMP A
PUMP B
Содержание ENO-30