15 |
P a g e
4-2. Power On
Turn on the main power switch on the left side of main body.
The display will light up after the power
is turned on.
A start screen will appear before entering the operation screen. The degasser vacuum
pump will immediately start and continue to run on and off throughout operation to maintain a vacuum.
Start Screen
*The EP-700 stores the parameters from the last time the pump was used and loads those
upon startup.
Status Indicator
The EP-700 indicator communicates the functional state of the instrument by color, on/off, or blinking.
Indicator
Subject
Lighting Pattern
DEGASS
Vacuum pump
degasser is
running
Turn off
Normal
Red Blinks
Abnormal
*
PUMP A
PUMP B
Machine motion
of PUMP A and
PUMP B
Turn off
Operation Off
Green Light
Operation On
Green Blinks
Pump Off timer is in
operation
*DEGASS When the indicator blinks red, there is a problem inside the
degasser. Please contact Eicom or one of its distributors immediately.
Indicator Lights