Guidance and Manufacturer’s Declaration – Immunity
The RETevet™ device is intended for use in the electromagnetic environment specifi ed below. The customer or user of the
RETevet™ device should ensure that it is used in such an environment.
Immunity Test IEC 60601
Test Level
Compliance Level
Electromagnetic Environment – Guidance
ESD
IEC 61000-4-2
± 8 kV Contact
± 15 kV Air
± 8 kV Contact
± 15 kV Air
Floors should be wood, concrete or
ceramic tile. If fl oors are synthetic, the r / h
should be at least 30%
EFT
IEC 61000-4-4
± 2 kV Mains
± 1 kV I / Os
± 2 kV Mains
± 1 kV I / Os
Mains power quality should be that of a
typical commercial, hospital, or home
environment
Surge
IEC 61000-4-5
± 1 kV Differential
± 2 kV Common
± 1 kV Differential
± 2 kV Common
Mains power quality should be that of a
typical commercial, hospital, or home
environment
Voltage
Dips / Dropout
IEC
61000-4-11
0 % U
T
; 0.5 cycle at 0°, 45°,
90°, 135°, 180°, 225°, 270°
and 315°
% U
T
; 1 cycle 70 %
U
T
; 25 / 30 cycles for 50 Hz
and 60 Hz, respectively
Single phase: at 0° 0 %
U
T
; 250 / 300 cycle for 50 Hz
and 60 Hz, respectively
Single phase: at 0°
0 % U
T
; 0.5 cycle at 0°, 45°,
90°, 135°, 180°, 225°, 270°
and 315°
% U
T
; 1 cycle 70 % U
T
;
25 / 30 cycles for 50 Hz and
60 Hz, respectively Single
phase: at 0° 0 % U
T
;
250 / 300 cycle for 50 Hz
and 60 Hz, respectively
Single phase: at 0°
Mains power quality should be that of a
typical commercial, hospital, or home
environment If the user of the RETevet™
requires continued operation during power
mains interruptions it is recommended
that the RETevet™ be powered from an
uninterruptible power supply or battery.
Power
Frequency
50 / 60 Hz
Magnetic Field
IEC 61000-4-8
30 A/m, 50 Hz or 60 Hz
30 A/m, 50 Hz or 60 Hz
Power frequency magnetic fi elds should be
that of a typical commercial, hospital, or
home environment.