![Eickemeyer 173610 Скачать руководство пользователя страница 15](http://html.mh-extra.com/html/eickemeyer/173610/173610_user-manual_2369194015.webp)
15
USER MANUAL
RETEVET™ PORTABLE ERG
T +49 7461 96 580 0 | F +49 7461 96 580 90 | [email protected] | www.eickemeyer.com
Rabbit, mini pig, photopic 2 step
Description
Eye
Flash Luminance Energy
(0.33, 0.33 White)
Background Luminance
(0.33, 0.33 White)
# Flashes
Light adapted 8.0
Right
8 cd·s/m² @ 2 Hz
30 cd/m²
20
Light adapted 8.0 fl icker
Right
8 cd·s/m² @ 28.3 Hz
30 cd/m²
141 – 424
Light adapted 8.0
Left
8 cd·s/m² @ 2 Hz
30 cd/m²
20
Light adapted 8.0 fl icker
Left
8 cd·s/m² @ 28.3 Hz
30 cd/m²
141 – 424
For the dark-adapted testing, dark adaption can be done overnight before the test starts or for at least 20 minutes before
testing. Use the red backlight option in
Settings
(see page 6) to start the test with a red screen. Oscillatory potentials are
automatically calculated from the DA 8.0 test, which is the 4
th
step in the protocol.
Rabbit, mini pig, scotopic 4 step
Description
Eye
Flash Luminance Energy
(0.33, 0.33 White)
Background Luminance
(0.33, 0.33 White)
# Flashes
Dark-adapted 0.6
Right
0.06 cd·s/m² @ 0.5 Hz
Off
3
Dark-adapted 8.0
Right
8 cd·s/m² @ 0.1 Hz
Off
3
Dark-adapted 25
Right
25 cd·s/m² @ 0.05 Hz
Off
3
Dark-adapted 0.06
Left
0.06 cd·s/m² @ 0.5 Hz
Off
3
Dark-adapted 8
Left
8 cd·s/m² @ 0.1 Hz
Off
3
Dark-adapted 25
Left
25 cd·s/m² @ 0.05 Hz
Off
3
For testing rats and mice, dimmer stimuli are typically used (Dembinska et al. 2001, Dembinska et al. 2002, Polosa et al. 2019).
Photopic and scotopic tests can be performed separately or together depending on the study design, there are separate
protocols for measuring just the animal’s right or left eye.
Rat, mouse photopic 2 step left eye
Description
Eye
Flash Luminance Energy
(0.33, 0.33 White)
Background Luminance
(0.33, 0.33 White)
# Flashes
Light adapted 8.0
Left
8.0 cd·s/m² @ 2 Hz
30 cd/m²
20
Light adapted 8.0 fl icker
Left
8.0 cd·s/m² @ 28.3 Hz
30 cd/m²
141 – 424
Rat, mouse photopic 2 step right eye
Description
Eye
Flash Luminance Energy
(0.33, 0.33 White)
Background Luminance
(0.33, 0.33 White)
# Flashes
Light adapted 8.0
Right
8.0 cd·s/m² @ 2 Hz
30 cd/m²
20
Light adapted 8.0 fl icker
Right
8.0 cd·s/m² @ 28.3 Hz
30 cd/m²
141 – 424
Overnight dark-adaptation is recommended in rodents. Oscillatory potentials are automatically calculated from the DA 8.0
test, which is the 3rd step in the protocol.