7. ADDITIONAL ACTIVITIES
Removing Old Results from the Device
The RETevet™ device can store up to 50 test results. You must remove old results to make room for new tests. There are three
ways to remove results.
Warning!
Results deleted on the device cannot be recovered. Save results you want to keep on a PC before deleting them off
the RETevet™ device.
Removing Selected Results from the Device
To remove individual results from the device, follow these steps:
1. Make sure that any results you want to keep have been copied to the PC.
2. Turn on the RETevet™ device.
3. Select Results.
4. Select the desired result to be erased.
5. Select Delete.
6. Select Yes.
Removing All Results from the Device
To remove all stored results from the device, follow these steps:
1. Make sure that any results you want to keep have been copied to the PC.
2. Turn on the RETevet™ device.
3. Select Settings then Memory.
4. Select Erase all test results.
5. Select Yes.
If during Step 4 you chose Erase everything, then the data storage area (including patient results and custom protocols) would
be deleted and reset to factory condition.
Removing Results Using the PC
To remove results from the device using a PC, follow these steps:
1. Place the RETevet™ device into the docking station.
2. Connect the USB cable.
3. Wait for the device to appear as an external drive on the PC.
4. Navigate to the Reports directory on the device.
5.
Make sure that any results you want to keep have been uploaded to the PC. Copy the fi les just as you would copy any fi le
from a thumb drive or other external device to a PC. If desired, also copy the corresponding raw data fi le (.rff) and XML fi le
(.rffx) from the / Data folder to archive the results in machine-readable formats for programmatic analysis.
6. Delete results from the Reports directory to remove them from the device. If you are saving results in multiple formats
(e. g., PDF and JPEG), all formats have to be deleted in order to remove the result from the device and make space for future
tests.
Removing and Replacing the Eyecup
To remove the eyecup, grasp the rubber nearest the silver bezel and pull gently.
To replace the eyecup, orient the eyecup so the slots in the white plastic on the eyecup are aligned with the bumps on the
device. Push gently until the eyecup clicks into the device. Replacement eye cups can be ordered from EICKEMEYER®.