![Eickemeyer 173610 Скачать руководство пользователя страница 11](http://html.mh-extra.com/html/eickemeyer/173610/173610_user-manual_2369194011.webp)
11
USER MANUAL
RETEVET™ PORTABLE ERG
T +49 7461 96 580 0 | F +49 7461 96 580 90 | [email protected] | www.eickemeyer.com
Protocol
ECVO 5-step single fl ash
is a modifi ed version of the ECVO 5-step protocol described above. The difference is while
the
ECVO 5-step
uses 4 fl ashes in the DA 0.01 and DA 3.0 steps in order to used averaging to improve signal quality, this
single
fl ash
version only uses 1 fl ash to minimize testing time. Replicates can be made in both protocols, and artefactual results
deleted. This protocol is recommended for NON-anesthetized or NON- sedated animals.
ECVO 5-step single fl ash
Description
Eye
Flash Luminance Energy
(0.33, 0.33 white)
Background Luminance
(0.33, 0.33 white)
# Flashes
Dark-adaptation timer
Both
Off
Off
Dark-adapted 0.01
Right
0.01 cd·s/m² @ 0.2 Hz
Off
1
Dark-adapted 3.0
Right
3 cd·s/m² @ 1 / 15 Hz
Off
1
Dark-adapted 10.0
Right
10 cd·s/m² @ 0.05 Hz
Off
1
Dark-adapted 0.01
Left
0.01 cd·s/m² @ 0.2 Hz
Off
1
Dark-adapted 3.0
Left
3 cd·s/m² @ 1 / 15 Hz
Off
1
Dark-adapted 10.0
Left
10 cd·s/m² @ 0.05 Hz
Off
1
Light adaptation timer
Right
Off
30 cd/m²
Light adapted 3.0
Right
3 cd·s/m² @ 2 Hz
30 cd/m²
20
Light adapted 3.0 fl icker
Right
3 cd·s/m² @ 28.3 Hz
30 cd/m²
141 – 424
Light adaptation timer
Left
Off
30 cd/m²
Light adapted 3.0
Left
3 cd·s/m² @ 2 Hz
30 cd/m²
20
Light adapted 3.0 fl icker
Left
3 cd·s/m² @ 28.3 Hz
30 cd/m²
141 – 424
Protocol
ECVO short, yes / no protocol
is a modifi ed version of the
ECVO 5-step single protocol
. Only the dark-adapted steps
are performed. This protocol is recommended for NON-anesthetized or NON-sedated animals when only rod- and mixed rod-
cone retinal function is required to be assessed.
ECVO short, yes / no protocol
Description
Eye
Flash Luminance Energy
(0.33, 0.33 white)
Background Luminance
(0.33, 0.33 white)
# Flashes
Dark-adaptation timer
Both
Off
Off
Dark-adapted 0.01
Right
0.01 cd·s/m² @ 0.2 Hz
Off
1
Dark-adapted 3.0
Right
3 cd·s/m² @ 1 / 15 Hz
Off
1
Dark-adapted 10.0
Right
10 cd·s/m² @ 0.05 Hz
Off
1
Dark-adapted 0.01
Left
0.01 cd·s/m² @ 0.2 Hz
Off
1
Dark-adapted 3.0
Left
3 cd·s/m² @ 1 / 15 Hz
Off
1
Dark-adapted 10.0
Left
10 cd·s/m² @ 0.05 Hz
Off
1