Updating Firmware
Periodically LKC publishes an update to the device fi rmware. Firmware updates are provided at no charge for the fi rst year,
afterward an annual support fee will be required for continued support and fi rmware updates.
You must download the update to the PC fi rst, connect the RETevet™ device to the PC, and carry out the fi rmware update
process. Follow these steps:
1.
Download the fi rmware update to the PC (follow instructions in the fi rmware update notice to fi nd and download the
update).
2. Connect the USB cable to the PC.
3. Place the device into the docking station.
4. Wait for the device to appear as an external drive on the PC.
5.
Copy the fi rmware update fi le from the directory on the PC to the Firmware directory on the device.
6. Eject the external drive that represents the device from the PC.
7. Remove the device from the docking station.
8. Select
Settings
, then
System
, then
Change settings
, then
Update Firmware
.
9.
Select the fi rmware update you want.
10. Select
Next
.
11.
Wait while the fi rmware is updated.
12.
After the fi rmware update completes, the device will restart automatically.
If the RETevet™ device fails during the fi rmware update, verify that the fi rmware update fi le was downloaded and copied to
the device correctly by repeating steps 5 through 12.