GeoScan User Manual | 8100550-03
•
2018-10-26 |
FOR SCIENTIFIC RESEARCH PURPOSES
109
Electromagnetic Immunity
Table A-2.
Electromagnetic compatibility (EMC) immunity guidelines and declarations for GeoScan
systems
GeoScan systems are intended for use in the electromagnetic environment specified
below. The customer or user of a GeoScan system should ensure that it is used in such an
environment.
Immunity Test
IEC 60601 Test
Level
Compliance
Level
Electromagnetic
Environment—Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±
6 kV contact
±
8 kV air
±
6 kV contact
±
8 kV air
Floors should be wood, concrete, or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±
2 kV for power-
supply lines
±
1 kV for
input/output
lines
±
2 kV for power-
supply lines
±
1 kV for
input/output
lines
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
±
1 kV differential
mode
±
2 kV common
mode
±
1 kV differential
mode
±
2 kV common
mode
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage dips, short
interruptions, and
voltage variations
on power-supply
input lines
IEC 61000-4-11
< 5% U
T
(> 95% dip
in U
T
) for
0.5 cycle
40% U
T
(60% dip in
U
T
) for 5 cycles
70% U
T
(30% dip in
U
T
) for 25 cycles
< 5% U
T
(> 95% dip
in U
T
) for 5 sec.
< 5% U
T
(> 95% dip
in U
T
) for
0.5 cycle
40% U
T
(60% dip in
U
T
) for 5 cycles
70% U
T
(30% dip in
U
T
) for 25 cycles
< 5% U
T
(> 95% dip
in U
T
) for 5 sec.
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the GeoScan system requires
continued operation during
power mains interruptions, it is
recommended that the GeoScan
be powered from an
uninterruptible power supply or
a battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical
commercial or hospital
environment.
Note:
U
T
is the AC mains voltage before application of the test level.
Содержание Geoscan
Страница 8: ...8 FOR SCIENTIFIC RESEARCH PURPOSES 8100550 03 2018 10 26 GeoScan User Manual ...
Страница 16: ...16 FOR SCIENTIFIC RESEARCH PURPOSES 8100550 03 2018 10 26 GeoScan User Manual ...
Страница 42: ...42 FOR SCIENTIFIC RESEARCH PURPOSES 8100550 03 2018 10 26 GeoScan User Manual ...
Страница 48: ...48 FOR SCIENTIFIC RESEARCH PURPOSES 8100550 03 2018 10 26 GeoScan User Manual ...
Страница 66: ...66 FOR SCIENTIFIC RESEARCH PURPOSES 8100550 03 2018 10 26 GeoScan User Manual ...
Страница 100: ...100 FOR SCIENTIFIC RESEARCH PURPOSES 8100550 03 2018 10 26 GeoScan User Manual ...
Страница 106: ...106 FOR SCIENTIFIC RESEARCH PURPOSES 8100550 03 2018 10 26 GeoScan User Manual ...
Страница 114: ...114 FOR SCIENTIFIC RESEARCH PURPOSES 8100550 03 2018 10 26 GeoScan User Manual ...
Страница 115: ......