Page 19
Edwards and the Edwards logo are trademarks of Edwards Limited.
Technical data
B800-00-880 Issue A
2.7
Electrical data
nEXT pumps can be driven either by your own system or by the Edwards TIC Turbo Instrument Controller or TIC Turbo
Controller.
If you wish to use your own system, the size of the power supply required depends on your application. The power
limit setting determines how quickly you can ramp up the pump and dictates the size of power supply you will need.
If you have serial communications capability, or have access to an Edwards TIC, you can select the power limit setting
of the nEXT pump. Refer to
for the maximum power limit settings for nEXT pumps. If your application
requires rapid cycling of the pump, you can achieve faster ramp times if you use a power supply that delivers higher
current, up to a maximum in accordance with
If you do not have the facility to adjust the power limit setting, you must use a power supply capable of delivering
enough current to meet the Edwards factory default power limit setting, shown in
If you intend to drive the nEXT pump using an Edwards TIC, be aware that there are several variants. Contact Edwards
to determine which is most suitable for your application.
2.8
Logic interface connector
nEXT pumps have a 15-way logic interface connector on the end of the logic interface cable. The logic interface
connector can be plugged directly into the Edwards TIC Turbo Instrument Controller or TIC Turbo Controller. You
must use a suitable connector mating half (not supplied) to connect the nEXT pump to your own equipment. Refer
to
for the connector mating half type, to
for Logic Interface connector pins for the electrical
connections and to Figure 7 for the connector interface circuit diagrams.
Table 11 - Logic interface technical data
Logic interface item
Connector
*
15-way D-type male
nEXT pumps electrical supply:
Allowable voltage range
(including any ripple)
24 - 48 V d.c. +5%, -10%
(21.6 to 50.4 V d.c.)
Maximum voltage ripple
0.5 V r.m.s.
Fuse rating
10A for 24Vdc supply
5A to 10A for 28Vdc supply
Type 'T' IEC approved or
Time delay fuse UL/CSA approved
Factory default setting
160 W
Maximum power limit setting
200 W
Minimum power limit setting
Accuracy of power regulation
50W
+/- 10W
Start, standby and serial enable control inputs:
Enabled control voltage: low (close)
0 to 0.8 V d.c. (I
out
= 0.55 mA nominal)
Disabled control voltage: high (open)
4 to 26.4 V d.c. (Internal pull up to 6.35 V nominal)
Analogue output:
Output voltage
0 to 10 V d.c. (directly proportional to measured
parameter)
Motor speed: 0 - 1000 Hz (0-100%)
Input power: 0 - 200 W
Motor temperature: 0 - 100 °C
Controller temperature: 0 - 100 °C
Output current
5 mA