![Edwards nEXT240 Скачать руководство пользователя страница 70](http://html1.mh-extra.com/html/edwards/next240/next240_instruction-manual_2366137070.webp)
© Edwards Limited 2009. All rights reserved.
Page 63
Edwards and the Edwards logo are trademarks of Edwards Limited.
Maintenance
B800-00-880 Issue A
5.5.1
Flashing Service Codes
Whenever a service is required the standard once per revolution flash on the yellow status LED is replaced with a
service flash code. The Service flash codes are specified in Table 23.
Table 26 - Flashing service codes
5.5.2
Flashing error codes
Whenever a Fail condition becomes active the red Alarm LED will slow the Error Flash Codes that reflect all active
error conditions. The Fail status flash code positions, within the Alarm LED error flash sequence, are specified in
There is a sufficient off period between each subsequent cycle repetition to clearly mark the start of a new flash
sequence. The duration of a long flash (L) is equal to 3 times the duration of a short flash (0.5s).
Service flash code
Comments
Actions
LED on 1sec
LED off 1sec
Oil cartridge service recommended
It is recommended that the oil
cartridge is replaced. Refer to Section
7.2
LED on 3sec
LED off 1sec
Bearing and oil cartridge service
recommended
It is recommended that the bearing
and oil cartridge are replaced. Refer
to Section 7.2.
LED on 3sec
LED off 3sec
Pump service required
The turbo impeller has reached its
expected life. It is recommended that
the pump is returned to an Edwards
service centre for service. Refer to
Section 7.2
Table 27 - Flashing error codes
Error flash
position
Error flash
code
Comments
Actions
0
ssssss
The speed fell below 50% of full
rotational speed with the Timer
disabled.
Check whether the pump is too hot or
whether the inlet pressure is too high.
1
Lsssss
Controller internal software mismatch. Cycle the power to the pump and see
whether the error code appears again.
If it does, contact your supplier or
Edwards.
2
sLssss
Controller failed internal configuration
and calibration operation.
Cycle the power to the pump and see
whether the error code appears again.
If it does, contact your supplier or
Edwards.
3
ssLsss
Failure to reach or maintain half full
speed within the timer setting value.
Check whether the pump is too hot or
whether the inlet pressure is too high.
4
sssLss
Overspeed or Overcurrent trip
activated, or other hardware fault.
Cycle the power to the pump and see
whether the error code appears again.
If it does, contact your supplier or
Edwards.
5
ssssLs
Pump internal measurement system
disconnected or damaged.
Cycle the power to the pump and see
whether the error code appears again.
If it does, contact your supplier or
Edwards.