Installation
B800-00-880 Issue A
3.7.5
Command set
shows a summary of the full set of commands available for controlling and monitoring the nEXT pump.
shows the abbreviations that are used to define commands in the following sections and
shows the
error codes that might be returned.
Table 17 - Summary of commands that can be sent to the nEXT pump
Object name
Command
Parameter
range
Factory
setting
Data
type
Units
Comments
Node
!S850
?S850
0.99
-
decimal
address Multi-drop address
0 = disable multi-drop address
99 = wildcard
Pump type
?S851
7;
10;
4
-
string
string
string
chars
chars
chars
Pump type
DSP software version number
(D396496XXX for nEXT)
Full speed RPS (1000 for nEXT)
Pump control
!C852
0
1
-
decimal
-
Stop the pump
Start the pump
?V852
0.1800;
32-bits
-
decimal
hex
RPS
flags
Measured motor speed
System status word
Vent options
!S853
?S853
0
1
2
3
4
5
6
7
8
0
decimal
Hard vent only when <50% speed
Controlled vent if >50% speed
or
hard vent if <50% speed
Hard vent if stop
or
hard vent if fail
and
<50% speed
Hard vent if stop
or
controlled vent
if fail
and
>50% speed
or
hard vent if
fail
and
<50% speed
Hard vent if fail
or
hard vent if stop
and
<50% speed
Hard vent if fail
or
controlled vent if
stop
and
>50% speed
or
hard vent if
stop
and
<50% speed
Hard vent if stop
or
fail
Same as option 6
Vent = Permanently Enabled (Fan)
Timer setting
!S854
?S854
1.30
8
decimal
minutes Timeout period for both initial ramp
up and if speed drops below 50%
Power limit setting
!S855
?S855
50.200
160
decimal
watts
Link power maximum
Normal speed
setting
!S856
?S856
50.100
80
decimal
%
Normal speed as a percentage of full
speed
Standby speed
setting
!S857
?S857
55.100
70
decimal
%
Standby speed as a percentage of
full speed
Temperature
readings
?V859
0.100;
0.100
-
decimal
decimal
°C
°C
Measured motor temperature
Measured controller temperature
Link parameter
readings
?V860
0.500;
0.300;
0.15000
-
decimal
decimal
decimal
0.1volts
0.1amps
0.1watts
Measured link voltage
Measured link current
Measured link power
Factory settings
!S867
1
-
-
-
Reset all configuration options and
parameters to the factory settings
PIC software
version
?S868
10
-
string
chars
Boot loader software version number
(D374796XXX)