![Edwards nEXT240 Скачать руководство пользователя страница 27](http://html1.mh-extra.com/html/edwards/next240/next240_instruction-manual_2366137027.webp)
B800-00-880 Issue A
Page 20
© Edwards Limited 2009. All rights reserved.
Edwards and the Edwards logo are trademarks of Edwards Limited.
Technical data
2.9
Controller connector socket
The nEXT pump has a 3-way Controller Connector Socket in the side of the Controller. When you receive the pump,
this connector will be concealed by a black protective cover. If you intend to use the connector, this cover should
be removed by levering with a small screwdriver. The mating plug for this connector is supplied with the pump.
The connector is intended to drive a vent valve or fan connected to the three pins. The connector mating plug is
shown in
, with the polarity of the pins marked when the vent valve/fan is energised.
NORMAL status output:
Type
< Normal speed (default 80%)
Off (2.2 k
internal pull up to 12 V d.c.)
Normal speed
On (<0.8 V d.c. sinking 20 mA)
Current Rating
Voltage Rating
20mA to 0V
28.8Vdc maximum external pull up voltage
FAIL status output:
Type
Open collector transistor
Fail
Off (3.3 k
pull up to 12 V d.c.)
OK
On (<0.1 V d.c. sinking 1.7 mA,
<0.8 V d.c. sinking 20 mA)
Current Rating
Voltage Rating
20mA to 0V
28.8Vdc maximum external pull up voltage
*
Mating half of connector not supplied.
Table 12 - Logic interface connector pins
Pin Number Signal
Polarity Use
2
0 V Control reference
-
0 V reference for all control and status signals below
3
START/STOP control input
-
Connect to Pin 2 to start pump
4
STANDBY control input /
Serial RX/RS485A-
-
Connect to Pin 2 to enable standby speed
5
Serial enable
-
Connect to Pin 2 to enable serial interface mode
7
FAIL / Serial TX/RS485 B+
-
Logic high when fail condition exists in parallel mode
9
Analogue output
Positive 0 - 10 V output proportional to measured output
10
Chassis / Screen
-
Screen
12
Chassis / Screen
-
-
15
NORMAL status output
-
Logic low when pump rotational speed is at normal speed or
above
8, 13, 14
Electrical supply: 0 V
-
1, 6, 11
Electrical supply: 24 V
Positive
Table 11 - Logic interface technical data (continued)