Object name
Command Parameter
range
Factory
setting
Data type
Units
Comments
Analogue sig-
nal options
!S871
0
0
decimal
-
Analogue output = meas-
ured speed
?S871
1
-
Analogue output = meas-
ured power
2
Analogue output = meas-
ured motor temp.
3
Analogue output = meas-
ured control temp.
4
Analogue output = meas-
ured rotor temp.
Electronic
braking op-
tions
!S872
0
0
decimal
Electronic braking = disa-
bled
?S872
1
Electronic braking = ena-
bled. Refer to
on page 14.
Close vent
valve
!C875
1
-
decimal
-
Closes the vent valve for
delayed start and over-
rides the current vent op-
tion. There is no open
vent valve command but
the stop command (!
C852 0) will clear the
override.
Valve 1 type
!S877
0
0
decimal
-
Normally Open Valve
?S877
1
Normally Closed Valve
(power cycle is required)
Valve 2 type
!S878
0
0
decimal
-
Normally Open Valve
?S878
1
Normally Closed Valve
(power cycle is required)
Service status
?V881
32 bits
-
hex
flags
Service status word
Controller run
time
?V882
0..999999
-
decimal
hours
Hours run by controller
0..999999
decimal
hours
Hours until controller
service due
Pump run
time
?V883
0..999999
-
decimal
hours
Hours run by pump
0..999999
decimal
hours
Hours until pump service
due
Pump cycles
?V884
0..65535
-
decimal
cycles
Cycles run by pump
0..65535
decimal
cycles
Cycles until pump service
due
Bearing run
time
?V885
0..999999
-
decimal
hours
Hours run by bearing
0..999999
decimal
hours
Hours until bearing serv-
ice due
07/2021 - ©Edwards Limited
Page 54
B8G000880_E
B8G000880_E - Installation