Page 20
-Select the line
„1st. Soot meas.“
.
-Press
<Enter>
to start the measurement. The analyzer will pull 1.63
liters of sample gas.
-Release the filter paper from the probe slot.
10. Soot test (continued)
-Compare the greyness to your soot test chart (0-9).
-Input the result using the numerical keys and press
<Enter>
.
-Repeat this procedure until all 3 soot tests are completed. The av-
erage value will be calculated and automatically stored.
-Go back to
Gas Analysis
and press
F2
to print off soot test results.
-Go to
Adjustments / Internal / Probe heating / F4
to turn off the
probe heater.
You do not need to input Boiler temp. or oil trace if not needed.
11. Averaging
The averaging feature allows you to collect snapshots in a timed
interval and calculates the average over a set time period. For in-
stance, if you want to take 1 measurement per minute over a 5 mi-
nute period, you can set up the Averaging test to automate this.
1. Select
Averaging
within the main menu
2. Select
Meas. Time
Input total measuring time in minutes or
seconds (for seconds, press decimal 1st)
3. Select
Scanning
Input interval time (for seconds, press
decimal 1st)
For example: If the
Meas. Time
is 5 minutes and the
Scanning
time
is 60 seconds or 1 minute, the mean value will take into account 5
scans and find the average.
Start measurement
Measurement time
Scanning
Printer
Store
Mean values
Select
:
Let the probe cool down before putting
back into the carrying case.