Page 19
Now go to
Adjustments
->
Units
and press up/down or right/left to
change the unit to
%O2 undiluted
. Your units have been changed
and you can return to Gas Analysis to view the corrected readings.
9. Draft/Pressure
A trend indication for the draft/pressure conditions in the exhaust
channel can be viewed during the gas analysis. However, the
draft/pressure value will not be on the printout until you zero the sen-
sor. To record a pressure measurement, follow these steps:
While in Gas Analysis, scroll through the pages until you get to this
one shown below. The pump will turn off.
Disconnect the draft hose from the in-
strument and press
F4
to zero the pres-
sure sensor. The sensor is now cali-
brated. Reattach the draft hose, wait for
the reading to appear, and press the
Save button
(number 6) to record the
measurement. Now when you press
F2
(whether in this menu or within other
Gas Analysis pages), you will see the
draft print out along with the gas read-
ings.
10. Soot test
The
Soot test
allows you to enter up to
3 soot test measurements.
You must first turn on the probe heater so that the sample is main-
tained above dew point, preventing any condensation in the probe
which can affect the filter paper. The filter paper slot is heated up to
approx. 70 °C. Go to
Adjustments / Internal / Probe heating / F1.
-Return to
Soot test.
-Insert a filter paper in the paper slot.
Pressure
Draught
-0.12
hPa
Recorded value:
--.-- hPa
Soot test
Select :
Boiler temp. : 66°C
1.Soot meas. : -.-
2.Soot meas. : -.-
3.Soot meas. : -.-
Oil trace : ----