8.1. Gas analysis display screen (continued)
Pressing
<F4>
goes to “Edit” mode which allows the user to change
the line items on the display screen.
In “Edit” mode, the values on the screen can be changed to what the
user wants to see. To change these values, press F4 and you should
see the 1
st
line start blinking:
- select line you would like to change with Up/Down
– it will blink
- select parameter you would like to see with Right/Left
- repeat until the display screen shows what you want it to
- press F4 to accept changes
8.2 Inserting the probe
Connect the sampling tubing on the instrument to the sample gas
inlet. Position the sampling probe in the exhaust channel, so that the
thermocouple is fully surrounded with the gas (see below).
Core stream search
Gas stream
Thermocouple hook
Probe tip
O2 3.2 %
CO2 13.1 %
CO 0 ppm
Eff. 92.5 %
Losses 7.5 %
Exc. air 1.18
T.Gas 184 °C
T.Air 20 °C
BImSchV
To control for thermal drift, start the analyzer and let it
acclimate to ambient temperature for 15-30 minutes,
then restart the analyzer before testing.