Page 12
6. Data Logging onto Memory Card (continued)
Memory card must be 32MB to 2GB!
To log data using the memory card, follow these steps:
-insert the memory card
– on the main menu a new option “Data
processing” will appear
-go to Data Processing
– press enter
-scroll down to Datalogger
– press enter
-scroll to Save to MMC
– press enter
-using the numerical buttons, enter the data logging interval in
seconds (min 1 sec, max 255 sec)
– press enter
(i.e. 15 sec will give you 4 data points per minute)
-scroll to Datalogger
– press enter – press F1 to confirm
-a blinking disk will appear in the upper right corner indicating that
the analyzer is currently logging data
-go back to Gas Analysis to start the pump (if not already running)
and view the data
-when you are finished with your test, go back to the Datalogger
screen (ESC - Data processing
– Datalogger)
-go to Datalogger
– press enter – press F1 to confirm to stop the test
The data file will be saved as a csv file (J2KDL-00.csv). Take out the
memory card and insert into a computer to view in Excel. Edit the
spreadsheet as necessary. The data fields on the csv file are listed
at the end of this manual.
In addition to data logger recordings the data could be transferred
online with USB cable (
USB Driver
/
1200 Baud
/ Protocol
DAS
) to
the software
“DASNT”
. The software
“DASNT”
and the
USB Driver
are available free of charge from www.ecomusa.com.