Page 18
If the measurement values are stable, press the
Save Button
(num-
ber 6) to transfer the values in the internal memory or
F2
to print.
8.5. Printing
Press
F2
to print the values on the screen.
To use the
Printer
button
(number 6 on the keypad), you have to
press the
Save button
(number 7 on the keypad)
before
pressing
the
Printer button
.
Inserting Text onto the Printout
If you want to insert text onto the printout, press
Enter
while in Gas
Analysis, select
Insert Text
, press up/down to select the line, and
press
Enter
again to start entering text.
When done, press
F1
to accept the text, then press
ESC
until you
are back in Gas Analysis.
Each time you print from now on, the text will appear on the printout.
To delete the text, press
Enter
while in Gas Analysis, select
Insert
Text
, press up/down to select the line, and press
Enter
again to start
editing the text. Press
F4
to get the cursor up to the text line, scroll
right until the blinking cursor is beyond the last character, then press
F2
until all characters are cleared.
When done, press
F1
to accept the text, then press
ESC
until you
are back in Gas Analysis. No text will print out until the next time you
insert text.
8.6 O2 Correction/Reference O2
O2 correction is commonly used by regulatory agencies to normalize
CO and NOx values to a specific O2 percent (usually 3% or 15%).
This allows for an “apples to apples” comparison when looking at
emissions data from many different pieces of equipment with many
different excess air values.
Go to
Adjustments
->
Ref. O2
and enter the desired O2 correction
value using the numbered buttons.
Measurement stored in
intermediate memory