Note:
Side and top slots in front panel of the unit are designed for tubing and capillaries
accommodation. FEP tubing with outer diameter 1
/8” are fitted by pressing into
th
e slot. Capillaries and tubing with outer diameter 1/16” are fitted using short
FEP tubing with outer diameter 1/8”
and
inner diameter 1/6”, threaded on the
capillary. A tubing threaded this way cannot be pressed in and it is fitted by
inserting from the side into the slot.
4.2. Door Installation
In order to prevent damage of door hinges, the door is supplied apart. The door should be
slipped on the hinges in opened position, see pictures below.
Door mounting must be easy, no strength needed, otherwise the hinges may get damaged.
4.3. Connection to Power Supply and Connection of Communication
Cables
Plug the power supply cord into the instrument socket on the rear panel. Connect PC via
Ethernet or using RS232. If you are connecting whole system, connect in the same manner also
other units.
4.4. Start of the Unit
Use a switch on the rear panel of the detector for connecting to mains voltage, then switch
on the unit using push button on the front panel. The push button glows red, if power source is off
and green, if power source is on. If power supply is disconnected, the unit remembers last status of
this push button and after power supply reconnection it restores the last status. This can be used
for automatic start in the system with a common switch of power source for whole system of liquid
chromatograph (for example using a multiple socket-outlet with a switch).