15. T
ECHNICAL
P
ARAMETERS
Parameter
Value
Wavelength range
190-800 nm (190-600 nm)
Spectral half-width
6 nm
Accuracy of adjustment
±
1 nm
Reproducibility
± 0
.5 nm
Light source
Deuterium discharge lamp and tungsten lamp *
Noise (Test cell, 254nm, TC 2 s, 10 Hz)
±
3
·
10
-6
AU
Drift (Test cell, 254nm)
1·
10
-4
AU/hr.
Time constant
20
–
10 000 ms
Sampling rate
Up to 100 Hz
Digital output
1 V/AU
Analogue output
1x configurable
Wetted materials
Depend on a cell
–
fused silica, PTFE, stainless
steel, Vespel, PEEK
Communication
RS232, Ethernet (LAN)
Display, keypad
VFD 140x32 pixels, 10 pushbuttons
Power supply
100-240 V 50/60 Hz 110 VA
Dimensions ( w x h x d )
280 x 135 x 498 mm
Weight
10.2 kg
Operational environment conditions
Indoor use only.
Altitude: up to 2000 m
Temperature: 5-
40 °C
Humidity: maximum relative humidity 80 % for
temperatures up to
31 °C
decreasing linearly to
50 % rel. humidity at
40°C.
Voltage fluctuations: up to
±10
% of nominal
voltage.
Overvoltage category II.
Pollution degree 2.
* Tungsten lamp only in version up to 800 nm.