SEFELEC1500M Instructions
145
V1.06
15.1.4.
DEVICE-DEPENDENT COMMANDS
MEG (MEGohmeter)
Activates the megohmeter function. Must be sent from the starting pattern.
PIC (PICOammeter)
Activates the current measurement function. Must be sent from the starting pattern.
SEQ (SEQuence)
Activates the sequence function. Must be sent from the starting pattern.
CONF (CONFig)
Activates the configuration function. Must be sent from the starting pattern.
PAR (PARameter) <NR1>
Selects the parameter set for the active function. E.g.:
MEG:PAR 1
DCV (DCVoltage) <NR1>
Changes the value of the test voltage of the current parameter set for the active func-
tion and switches to DC voltage if the option is present in strength.
E.g.: DCV 500
HTIM (HTIMe) <NR1>
or <NR2>
Changes the value of the maintaining time in seconds of the current parameter set for
the active function. Values between 0.1 and 9,999
E.g.: HTIM 3
RTIM (RTIMe) <NR1> or <NR2>
Changes the value of the rising time in seconds of the current parameter set for the ac-
tive function. Values between 0 and 9,999
E.g.: RTIM 10
FTIM (FTIMe) <NR1> or <NR2>
Changes the value of the falling time in seconds of the current parameter set for the ac-
tive function. Values between 0 and 9,999
E.g.: FTIM 5
HLIM (HLIMIt) <NR3>
Changes the value of the upper threshold of the current parameter set for the active
function.
E.g.:
MEG:HLIM 2.0E+6
(new value in Ohms)
PIC:HLIM 1.45E-6
(new value in Amperes)
The unit is automatically expressed in Volts, Ohms or Amperes and depends on the
function.
CBHLIM <NR3>
Same as HLIM but resetting of the lower threshold before loading the value of the upper
threshold
.
Содержание SEFELEC 1500M
Страница 1: ...SEFELEC 1500M PENT6570 version 1 06 January 2020 User Manual...
Страница 34: ...SEFELEC1500M Instructions 33 V1 06 Continue in the same way to set the time and validate with OK...
Страница 122: ...SEFELEC1500M Instructions 121 V1 06 Diagram no 2 Running an insulation test Memory 5 Test result fail...
Страница 123: ...SEFELEC1500M Instructions 122 V1 06 Diagram no 3 Execution of a sequence Memory 3 Result pass PLC SYSTEM DEVICE...
Страница 137: ...SEFELEC1500M Instructions 136 V1 06 In the last two windows opened click on OK 14 5 CONFIGURING THE ETHERNET CONNECTION...
Страница 163: ...SEFELEC1500M Instructions 162 V1 06 20 EU DECLARATION OF CONFORMITY...
Страница 164: ...SEFELEC1500M Instructions 163 V1 06...