8.3.
ERROR MESSAGES:
➢
"SAFETY LOOP OPEN"
: the safety loop circuit is not closed, there is no connection be-
tween points 1-9 or 2-10 of the socket C5 located on the rear panel of the device. Press
OK to clear the message
➢
"OVER RANGE"
: the insulation resistance of the sample being tested is greater
than the measuring possibilities of the device
(> 2,000 TΩ) or of the measuring
range in manual mode.
➢
"UNDER RANGE"
: the insulation resistance of the sample being tested is less than the
measuring possibilities of the device (<
100 Ω) or of the measuring range in manual
mode.
➢
"CHARGING....
": the measurement voltage has not reached its final value. The voltage
generator charges the capacitor of the sample being tested. This message must appear
in a fleeting manner when switching to Measurement mode. If the message persists, this
means that the measurement voltage requested cannot be supplied to the resistance
measured.
Check with respect to the characteristics of the voltage generator.
➢
"VOLTAGE ERROR
": when a measurement time is defined, if during the test the mes-
sage
"CHARGING
…"
is constantly displayed, at the end of the measurement time the
test will be declared as FAIL and the
"VOLTAGE ERROR"
message will indicate the
cause of it.
If needed see the information on the message
"CHARGING
…"
IMPORTANT
The device is equipped with a thermal protection system. In case of pro-
longed use of the High Voltage generator on a short circuit, the protection
system will be activated.
In this case the message "
OVERHEAT
" will be displayed constantly on the
LCD screen.
Turn off the device and allow it to cool down for 10 minutes before using it
again.
Содержание SEFELEC 1500M
Страница 1: ...SEFELEC 1500M PENT6570 version 1 06 January 2020 User Manual...
Страница 34: ...SEFELEC1500M Instructions 33 V1 06 Continue in the same way to set the time and validate with OK...
Страница 122: ...SEFELEC1500M Instructions 121 V1 06 Diagram no 2 Running an insulation test Memory 5 Test result fail...
Страница 123: ...SEFELEC1500M Instructions 122 V1 06 Diagram no 3 Execution of a sequence Memory 3 Result pass PLC SYSTEM DEVICE...
Страница 137: ...SEFELEC1500M Instructions 136 V1 06 In the last two windows opened click on OK 14 5 CONFIGURING THE ETHERNET CONNECTION...
Страница 163: ...SEFELEC1500M Instructions 162 V1 06 20 EU DECLARATION OF CONFORMITY...
Страница 164: ...SEFELEC1500M Instructions 163 V1 06...