SEFELEC1500M Instructions
152
V1.06
The contribution of the cable to the measured leakage current decreases exponentially
with time. It is, therefore, important to specify the time taken for the test in order for the re-
sult to be significant. SEFELEC1500M devices can satisfy this requirement thanks to the
built-in timer, able to time from
0.1 s to nearly 3 hours
.
16.4.4.
TEST VOLTAGE FOR MEASUREMENT
Insulation resistance measurements are used to check that materials or equipment meet
defined standards; it is important to refer to these standards when choosing the test volt-
age. Standard test voltages are usually:
50, 100, 250 and 500
DC volts. If there is no rec-
ommended test voltage, choose a voltage of
100
volts for the measurements.
Where measurements are made on capacitive devices and when the insulation resistance
is to be measured at different voltages, always start from the lowest test voltage and in-
crease the voltage for subsequent measurements. Reducing the test voltage may give in-
consistent results.
Содержание SEFELEC 1500M
Страница 1: ...SEFELEC 1500M PENT6570 version 1 06 January 2020 User Manual...
Страница 34: ...SEFELEC1500M Instructions 33 V1 06 Continue in the same way to set the time and validate with OK...
Страница 122: ...SEFELEC1500M Instructions 121 V1 06 Diagram no 2 Running an insulation test Memory 5 Test result fail...
Страница 123: ...SEFELEC1500M Instructions 122 V1 06 Diagram no 3 Execution of a sequence Memory 3 Result pass PLC SYSTEM DEVICE...
Страница 137: ...SEFELEC1500M Instructions 136 V1 06 In the last two windows opened click on OK 14 5 CONFIGURING THE ETHERNET CONNECTION...
Страница 163: ...SEFELEC1500M Instructions 162 V1 06 20 EU DECLARATION OF CONFORMITY...
Страница 164: ...SEFELEC1500M Instructions 163 V1 06...