SEFELEC1500M Instructions
141
V1.06
The transmission parameters cannot be modified and are:
Parity:
without
Format:
8 bits
Stop bit:
1
To modify the transmission speed, go to the CONFIG menu, on the INTERFACE line and se-
lect RS232. In the Baud rate area select the speed by scrolling through the values between
1200 bauds to 115 200 bauds. Confirm by pressing the desired number.
A command intended for the device
must
end with the character:
LF
(hexadecimal
0A
,
decimal
10
).
Codes can be sent in upper or lower case. The maximum number of commands in the
same block cannot exceed 15 instructions.
Be careful not to associate standard com-
mands or requests with other types of commands.
The end of the execution of a block of commands by the device is reported via the emis-
sion of the
Xon
character (hexadecimal
11
, decimal
17
) and makes it possible to synchro-
nise the communication with the computer as the latter has to wait for the
Xon
character
before sending the following blocks.
The events that occur during the measurement can be reported to the computer through
emitting a
Z
character (format errors, end of test, open safety circuit, etc.). To activate this
function, the code
SRQ
must be sent after the device has initialised. When the
Z
character
is received, a special command allows the computer to accurately know the event (see
*STB?
and
*ESR?
commands).
In the case of incorrect syntax, the message
DIALOGUE ERROR: 1
is displayed. In the
case of a code that is out of context the message
DIALOGUE ERROR: 2
is displayed
(specific code for a function when we are not in this function, or numerical value out of
bounds).
Содержание SEFELEC 1500M
Страница 1: ...SEFELEC 1500M PENT6570 version 1 06 January 2020 User Manual...
Страница 34: ...SEFELEC1500M Instructions 33 V1 06 Continue in the same way to set the time and validate with OK...
Страница 122: ...SEFELEC1500M Instructions 121 V1 06 Diagram no 2 Running an insulation test Memory 5 Test result fail...
Страница 123: ...SEFELEC1500M Instructions 122 V1 06 Diagram no 3 Execution of a sequence Memory 3 Result pass PLC SYSTEM DEVICE...
Страница 137: ...SEFELEC1500M Instructions 136 V1 06 In the last two windows opened click on OK 14 5 CONFIGURING THE ETHERNET CONNECTION...
Страница 163: ...SEFELEC1500M Instructions 162 V1 06 20 EU DECLARATION OF CONFORMITY...
Страница 164: ...SEFELEC1500M Instructions 163 V1 06...