/ Programmable DC Electronic Load - User Manual
2.5.15. Qualification Test and Its Limit
Qualification Test (qUALI or quality) can be enabled or disabled.
2-5-15-1 Qualification Test Setting
When the Qualification Test is enabled, the channel is active, and the tested voltage, current
and power exceed the limits set for the test, the "OK" icon will flash, accompanied by a sounding
Buzzer. If the tested voltage, current and power are within the specified limits, the "OK" icon will
remain steady, and the Beep will remain silent.
When the Qualification Test is enabled, upper and lower limits for voltage, current, and
power can be edited.
Users can press the [ENT] button and use the arrow buttons and the knob to adjust the value.
2-5-15-2 Parameters Configurable When the Qualification Test Is Enabled
When the Qualification Test is disabled, these six values will appear as "NONE," indicating
that they cannot be edited.
2-5-15-3 When the Qualification Test Is Disabled, No Parameters Can Be Modified
2.5.16. System Information
System Information (InFo or Information) is for display only and cannot be edited. It shows
the current software version and serial number, with the software version displayed on the second
line and the serial number on the third line.
2-5-16-1 System Information
3. Basic Mode Operations
The electronic load features four basic operating modes:
Constant Current (CC), Constant Voltage (CV), Constant Resistance (CR), and Constant
Power (CP).