/ Programmable DC Electronic Load - User Manual
LIST test, the testing procedures and parameters for the device under test can be organized into a
list and executed in sequence. Specific setting parameters include: defining the number of steps,
step mode, loop switch, load mode for each step, load magnitude, delay time, comparison switch,
upper limit, and lower limit.
5.1. Main Display Interface
When the channel is deactivated, it can display basic electrical parameters (voltage, current,
power), or the 10 list settings (excluding delay settings) or test results from the previous LIST
setting test. Use the arrow buttons [▲ ▼◄ ►] to switch between the displayed interfaces.
Fig. 5-1-1 Basic LIST Electrical Parameters and LIST Test Results
Refer to section 5.3 for details on the LIST settings interface.
When the channel is activated, voltage will be displayed on the first line, with current,
resistance or current LIST steps on the second line and power on the third line. Users can use the
arrow buttons [▲ ▼◄ ►] to select whether to display the parameters alternatively or to display a
specific parameter permanently on the second line.
Fig. 5-1-2 LIST Test When the Channel Is Activated
5.2. Parameter Editing
Parameters cannot be edited while the channel is activated; editing is only possible when the
channel is deactivated.
Press the [ENT] button to enter the parameter setting interface and select parameters to be
edited by using the knob or arrow buttons. After the selection, press the [ENT] button again to edit
the selected parameter. The following parameters can be edited.
After setting parameters, press the [ESC] button repeatedly to return to the main display
interface.
Test Steps:
Set the number of test items, ranging from 1 to 10.