/ Programmable DC Electronic Load - User Manual
the selected parameter. The following parameters can be edited.
After setting the parameters, press the [ENT] button repeatedly to return to the main display
interface.
Discharge Modes
: In Battery mode, there are two discharge modes, CC (constant current
discharge) or Cr (constant resistance discharge).
Fig. 7-2-1 Discharge Modes
Cut-off Conditions:
In Battery mode, there are four cut-off conditions, voltage (voLt), time
(tIME), capacity (AH) and energy (WH).
Fig. 7-2-2 Cut-off Conditions
Cut-off Enable:
In the constant current (CC) discharge mode with voltage (voLt) as the cut-off condition,
users can select settings from 1 to 3, adjusting the discharge current and cut-off voltage within this
range.
In other scenarios, the setting is limited to 1.
Fig. 7-2-3 Cut-off Enable
Discharge Setting Value
: In the CC discharge mode, it is set to current; in the Cr discharge mode,
it is set to resistance.
Fig. 7-2-4 Discharge Setting Value
In the CC discharge mode with voLt as the cut-off condition, the number of discharge setting
values that can be configured is equal to the cut-off enable value.
Cut-off Voltage:
Set the cut-off value when voltage is selected as the cut-off condition.
In the CC discharge mode with voLt as the cut-off condition, the number of cut-off voltage
values that can be configured is equal to the cut-off enable value.