5-2-1 Test Steps
Step Mode:
Continuous/Trigger. Set the step mode for test items.
Continuous: Perform tasks sequentially and sustain the delay time in the current mode.
Trigger: Enter the next mode upon receiving a trigger signal.
5-2-2 Test Steps
Looping:
Enabled/Disabled. Set whether the entire test will be performed repeatedly.
When the looping mode is disabled, operation starts from the mode 1. Upon the completion of
the final mode, the channel is automatically deactivated, thus concluding the operation.
When the looping mode is enabled, operation starts from mode 1 and, upon the completion of
the final mode, loops back to mode 1 to start over again.
5-2-3 Looping Mode
Modes:
The following 6 modes are available.
CC (Constant Current), CV (Constant Voltage), CP (Constant Power), CR (Constant
Resistance), oPEn (Open Circuit) and SHort (Short Circuit).
5-2-4 Functional Modes
Setting Values:
Correspond to the test setting values for the respective test items.
In CC mode, current can be set; in CV mode, voltage can be set; in CP mode, power can be
set; in CR mode, resistance can be set.
In oPEn or SHort mode, settings are unavailable, with nonE being displayed.
5-2-4 Setting Values in Different Modes
Note:
When the mode is changed, the setting value will be limited to the maximum setting
within the current range. For instance, if the initial setting in CV mode is 100V, when the CV
mode is switched to the CC mode, the low current range will be capped at 3A, with the high
current range limited to the maximum current setting.
Delay
: Runtime setting for this specific test item.