/ Programmable DC Electronic Load - User Manual
capacity
Resolution
1mA
,
10mA
,
10m
Ω,
1
Ω
Range of measurement
Voltage
read-back
value
Range
0~19.999V,0 ~120.00V
0~19.999V,0~180.00V
Resolution
1mV,10mV
Accuracy
±(
0.05%+0.1%FS
)
Current
read-back
value
Range
0~3.000A,0
~
20.00A
0~3.000A,0~30.00A
Resolution
1mA,10mA
Accuracy
±(
0.05%+0.1%FS
)
±(
0.05%+0.1%FS
)
Power
read-back
value
Range
200W
Resolution
10mW
Accuracy
±(
0.1%+0.5%FS
)
Scope of protection
Over Voltage Protection(OV)
>
21V OR
>
125V
over voltage
protection
>
21V OR
>
185V
over voltage
protection
Over Current Protection(OC)
>
3.1 or
>
21
A over current
protection
>
3.1 or
>
31
A over current protection
Over Power Protection(OP)
210W
Over Temperature Protection
85
℃