Mechanical/pneumatic parts - Operation
6.1
Vortex cooler
•
The optional Vortex coolers create a positive pressure, cool air flow inside the RSU
and RCU to keep the ViscoIndicator running under optimal conditions
•
Each Vortex cooler should be fed clean filtered air at a minimum pressure of 30psi
from the Dynisco supplied pressure regulator/filter (Dynisco p/n 9967-3759)
•
Even when the ViscoIndicator isn't at temperature, it is preferred to keep the Vortex
coolers on with air flowing to maintain a positive pressure air flow inside the RSU
and RCU enclosures to limit the accumulation of dust/particulates in the enclosures
6.2
IRCU-STAND casters
•
After the connection is made between the VI system assembly and the process
connection, the IRCU-STAND casters should be locked in position if the VI system
isn't subject to significant shock loads.