4.4
Attach the HMI to the stand
The HMI consists of a capacitive touchpanel mounted into a metal box.
The touchpanel electronics must be mounted in a dry location that does not exceed 50C.
Item numbers refer to ViscoIndicator Stand Drawing (IRCU-STAND) in section 4.1.
4.4.1
Insert M5 T-Nuts
•
Position four M5 T-Nuts (item 17) into 2 slots towards the top end of item 3 (yellow
circles)
4.4.2
Place HMI on top of square post and open HMI box
•
Place the HMI on top of the square post (item 1) against item 3
•
Open the HMI box by unscrewing the 4 screws in the cover panel, being careful to
hold the box and cover panel in place.
USE CAUTION FRAGILE: The cover panel is electrically attached to the box via
a fragile wire so please be careful when handling it after unscrewing. Do not
pull or stretch the electrical cable.