2 MediTEST User Guide Page 12
-
2
SYSTEM CALIBRATION
12.2 RECALIBRATION SETTINGS
The parameters set the recalibration period and the recalibration reminder period.
Open the
SYSTEM SETTINGS
screen.
Touch
-
on, or select a parameter and a data entry screen will open (Fig 12.1).
Enter the required values via a keyboard or the increment / decrement controls (as described in Fig
12.1) and press
ENTER
to save (or
CANCEL
to abandon).
12.3 LOAD CELL SETTINGS
The Load Cell Settings parameters are critical to the system function. They should not be modified
(unless under the direction of Dynascan). They should be as is listed in section 5.3
Do not modify these settings, If this screen is accessed, press
CANCEL
to exit safely.
CAUTION
Specifies the number of days a calibration is valid
(before it expires).
Specifies the number of days before a calibration
expires that a reminder is posted.
Figure 12.1
Figure 12.2