Installation and Operation of X-STREAM WASH
16
Configuration Mode –
CnF
.
Enter the supervisor mode and press the
Up
or
Down
buttons until CnF. is displayed. Press the
TEMP
button
to select.
When prompted for a password (PSS
), use the
Up
and
Down
buttons to select 415. Press the
TEMP
button to enter in the value. Press the
TEMP
button
to toggle between the parameters that are available,
see parameter table below.
The parameter name and its value will be swapped in
and out of the display. Press the
Up
and
Down
button
to change the parameters value. Once a value has been
changed, press the
TEMP
button to lock in this value
and move on to the next parameter. If you do not press
the
TEMP
button within three seconds after changing
a parameters value, the new value will not be saved.
Exit the supervisor mode when done with the setup mode.
NOTE: See Supervisor Mode section on how to exit.
Configuration Mode Parameter Table
Parameter
Mnemonic
Description
Table 1 Default
Value (Celsius)
Table 2 Default
Value (Fahrenheit)
Units
P1
Input table
2
10
P2
Low scale range of sensors
0°C
32°F
Degrees
P3
High scale range of sensors
282°C
540°F
Degrees
P4
Output 1 control action
rEu
rEu
P5
Output 2 function
I
I
P6
Output 2 alarm type
H.A.
H.A.
P8
Alarm 1 standby function
OFF
OFF
P11
Auto-ranging display
OFF
OFF
P12
Filter sensor values
On
On
P22
Output 3 function
I
I
P23
Output 3 alarm type
H.A.
H.A.
P25
Alarm 2 standby function
OFF
OFF
P33
Access to OP.S mode
55
55
P35
Access to CnF. mode
415
415
P36
Access to I.CL mode
415
415
Loading Celsius or Fahrenheit Tables
The controller contains two default parameter tables.
Table 1 is used for Celsius operation, and Table 2 is
used for Fahrenheit.
To load one of the default tables, enter the configuration
mode (See instructions listed in the Configuration Mode
–
CnF section listed above). When CnF
is displayed,
press both the
Up
and
Down
buttons simultaneously.
The display will show
dL.F. Press the
Up
or
Down
buttons to display
dL.1
for table 1 (Celsius), or
dL.2
for table 2 (Fahrenheit).
Press the
TEMP
button to enter in your selection. The
display will show
Ld.t. This indicates that the loading
procedure has been initiated and after approximately
1 second, the loading will be finished and the display
will revert back to
CnF.
Exit the supervisor mode when done.
NOTE: See
Supervisor Mode section on how to exit.
Содержание X-STREAM WASH
Страница 27: ...27 Installation and Operation of X STREAM WASH K1000 208 240V 1 PHASE UNITS TO 12 2013...
Страница 28: ...Installation and Operation of X STREAM WASH 28 K1000 208 240V 3 PHASE UNITS AFTER 12 2013...
Страница 29: ...29 Installation and Operation of X STREAM WASH K1000 208 240V 3 PHASE UNITS TO 12 2013...
Страница 30: ...Installation and Operation of X STREAM WASH 30 K1000 440 480V 3 PHASE UNITS AFTER 12 2013...
Страница 31: ...31 Installation and Operation of X STREAM WASH K1000 440 480V 3 PHASE UNITS TO 12 2013...
Страница 34: ...Installation and Operation of X STREAM WASH 34 K5000 208 240V 3 PHASE W SANITIZER...
Страница 35: ...35 Installation and Operation of X STREAM WASH K5000 420 480V 3 PHASE W SANITIZER...