4-10
Communications
Communications
Communication
interface setup
Communications Settings is where instrument-specific information is entered to allow
the PowerVisa to communicate to external devices using the optional Isolated
Communications Module assemblies or modem (see related information on Chapter 9
Downloading Events).
Action...
Result...
STEP 1: From the Instrument
Settings Setup Menu screen, press
Communications.
• Press
Enable Communications
to
activate/deactivate the
communications port. External
communication devices connect to
the PowerVisa mainframe via the
optical serial data port. The optical
serial data port connector is
located on the left side of the
instrument.
• Press to enter the communications
Device Address
from 01 to 32 for
use with the optional Fiber Optic
to Ethernet adapter.
• Select the corresponding
Baud
Rate
to set connection speed.
Default baud rate is 11520.
• Press
External Modem
Connected
if you want to activate
communications via an external
modem.
Press
Change
to modify
the modem initialization string.
The string must begin with “AT”.
Dranetz-BMI recommends using
the same brand and model of
modem at both ends of the
communications link. See Chapter
9 Downloading Events for
instructions on modem
connections.
• Press
IP
to configure your
Ethernet settings and allow the
PowerVisa to operate on the
network. Proceed to Step 2 on
page 4-11.
• Press
OK
to accept changes in
communications settings.
• Press
Cancel
to discard changes in
communications settings and
return to Instrument Settings
menu.
MARK159
NOTE: For instructions on how to physically
connect to NodeLink using the external
communications devices, see Chapter 9
Downloading Events.
Содержание PowerVisa
Страница 1: ...PowerVisaTM USER S GUIDE DRANETZ BMI 1000 New Durham Road Edison New Jersey 08818 4019...
Страница 13: ...xiii Dranetz BMI PowerVisaTM...
Страница 14: ......
Страница 78: ...4 16...
Страница 136: ...6 22...
Страница 158: ...7 22...
Страница 204: ...A 10...
Страница 221: ...C 5 APPENDIX C Battery Specifications and Replacement Procedure Battery removal diagram...
Страница 222: ...C 6...
Страница 224: ...D 2...
Страница 248: ...F 4...
Страница 251: ...LEVEL 100 1 Start Menu Automatic Setup MARK101a MARK104 MARK103a G 3 MARK100 To Level 001 VISA123 MARK104a...
Страница 257: ...LEVEL 150 Instrument Settings MARK151 VISA152 PVISA155 MARK156 MARK157 MARK158 MARK160 MARK159 G 7...
Страница 259: ...MARK201 LEVEL 200 Scope Mode MARK202 MARK205 G 9...
Страница 260: ...MARK301PG LEVEL 300 1 Meter Mode Standard MARK302PG MARK303PG MARK305PG MARK307PG MARK306PG MARK304PG G 10...
Страница 261: ...LEVEL 300 2 Meter Mode Distortion MARK315PG MARK313PG MARK314PG MARK312PG MARK316PG MARK311PG G 11...
Страница 268: ...Level 112_601 Advanced Options Journal Interval MARK112_401 MARK112_402 MARK112_403 MARK112_404 MARK112_405 MARK112 G 18...
Страница 269: ...MARK112b LEVEL 112_701 Advanced Options EN50160 Options MARK112c G 19...
Страница 272: ...MARK431 LEVEL 430 Trend MARK432 MARK437b MARK433PG MARK436 MARK434 MARK435 MARK438 MARK440 MARK441 MARK439 MARK437 G 22...