Interlock system
The interlock system ensures that on anaesthetic
inhalation machines with two Vapors 19.n* only one
Vapor can be used at any time, and that the other Vapor
is blocked.
Interlock 2
For two Vapors 19.3 with plug-in systems.
Illustration: left, Vapor blocked,
right, Vapor operational.
Interlock 1
For two permanently fixed Vapors 19.2.
Illustration: right, Vapor blocked,
left, Vapor operational.
_____________
*
Vapor 19.1 is not suitable for Interlock systems 1 or 2.
However, see "Model Summary", page 42 for connection options.
Vapor 19.1 can be modified by DrägerService so that it can be
connected to Interlock 1 or 2.
e
i
0
Vol. %
Vol. %
7
Method of Operation
Interlock system
i
e
0
Vol. %
Vol. %