Vapor with key-indexed filling system
●
Keep Vapor vertical.
1 Switch off Vapor = set handwheel to »0« –
»0« button engaged.
When filling during operation:
●
The fresh gas flow can remain as set.
●
After handwheel has been set to »0« wait for
5 seconds for pressure to equalise.
If the filling system is opened straight away, the liquid
anaesthetic agent will bubble out because the
vaporising chamber is pressurised when the Vapor is
switched on and in operation.
●
Use bottle with indexed collar.
●
Screw a suitable adaptor firmly onto bottle.
2 Turn fill/drain valve to »0« position.
3 Undo locking screw to the stop, as the seal might
otherwise be damaged when the adaptor is inserted.*
4 Pull out sealing block.
______________
*
The overfill safety mechanism of the key-indexed filling system
requires a flexible seal which, however, is sensitive to mechanical
stress. Its service life can be extended if the locking screw, when
removed, is always unscrewed right to the stop. The seal is a wear
part and may only be replaced by duly qualified specialists.